Experiment
Title
Co K edge XAS transmission and XAS fluorescence and XAS HERFD of Co oxides and salts for the study of Co doping in maghemite nanoparticles
DOI
10.26302/SSHADE/EXPERIMENT_IM_20120926_001
Data reference
Vichery, Charlotte; Maurin, Isabelle (2012): Co K edge XAS transmission and XAS fluorescence and XAS HERFD of Co oxides and salts for the study of Co doping in maghemite nanoparticles. SSHADE/FAME (OSUG Data Center). Dataset/Spectral Data. https://doi.org/10.26302/SSHADE/EXPERIMENT_IM_20120926_001
Publications
Type(s)
laboratory measurement
Description
Varying the Co environment and adapting the observation mode as a function of the Co content
Number of spectra
8
Variable type(s)
  • observation mode
  • sample composition
Instrument
X-ray Absorption Spectroscopy measured in transmission mode
Sample holder
simple holder
Standard medium
unknown
Observation mode
spectrum
Spectral range type(s)
hard X
Valid spectral range(s)
Min - Max ($eV$) Sampling ($eV$) Resolution ($eV$) Position accuracy ($eV$) Absorption edge
#1 7550.0 - 8451.0 0.4 $Co$ (K: 7.7089 $keV$)
Comments
The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the absorption edge of a reference: Co metallic foil (first maximum of the derivative set at 7709 eV).
Observation geometry
direct
Observation mode
fixed angles
Observation mode
single spot
Instrument
X-ray Absorption Spectroscopy measured in fluorescence mode
Sample holder
simple holder
Standard medium
unknown
Observation mode
spectrum
Spectral range type(s)
hard X
Valid spectral range(s)
Min - Max ($eV$) Sampling ($eV$) Resolution ($eV$) Position accuracy ($eV$) Absorption edge
#1 7550.0 - 8451.0 0.4 $Co$ (K: 7.7089 $keV$)
Comments
The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the absorption edge of a reference: Co metallic foil (first maximum of the derivative set at 7709 eV).
Observation geometry
directional
Observation mode
fixed angles
Comments
Emitted x-rays are collected in a solid angle directed at 90 degrees from the incident beam.
Observation mode
single spot
Instrument
High Energy Resolution Fluorescence Detection X-ray Absorption Spectroscopy
Sample holder
simple holder
Standard medium
unknown
Observation mode
spectrum
Spectral range type(s)
hard X
Valid spectral range(s)
Min - Max ($eV$) Sampling ($eV$) Resolution ($eV$) Position accuracy ($eV$) Absorption edge
#1 7550.0 - 8088.0 0.4 $Co$ (K: 7.7089 $keV$)
Comments
The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the absorption edge of a reference: Co metallic foil (first maximum of the derivative set at 7709 eV). The resolution of the spectrometer analyser was 5.2 eV (FWHM of the elastic peak).
Observation geometry
directional
Observation mode
fixed angles
Comments
Emitted x-rays are collected in a solid angle directed at 90 degrees from the incident beam.
Observation mode
single spot
Date begin
2012-09-26

Versions

Release date
2019-09-02 06:51:12 UTC+0000
Version (Date)
#1 (2019-09-02 06:51:12 UTC+0000, Updated: 2019-09-02 06:49:05 UTC+0000)