Experiment
Title
Br K edge XAS transmission and HERFD of sodium bromide anydrous salt at ambient conditions
DOI
10.26302/SSHADE/EXPERIMENT_OP_20180115_002
Data reference
Proux, Olivier; Irar, Mohammed (2015): Br K edge XAS transmission and HERFD of sodium bromide anydrous salt at ambient conditions. SSHADE/FAME (OSUG Data Center). Dataset/Spectral Data. https://doi.org/10.26302/SSHADE/EXPERIMENT_OP_20180115_002
Publications
Type(s)
laboratory measurement
Number of spectra
2
Variable type(s)
  • observation mode
Instrument
X-ray Absorption Spectroscopy measured in transmission mode
Standard medium
unknown
Observation mode
spectrum
Spectral range type(s)
hard X
Valid spectral range(s)
Min - Max (eV) Sampling (eV) Resolution (eV) Position accuracy (eV) Absorption edge
#1 13351.0 - 13715.0 0.7 Br (K: 14.3256 keV)
Comments
The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the L2 spectrum of a reference Pt metallic foil (first maximum of the derivative set at 13273 eV).
Observation mode
single spot
Instrument
High Energy Resolution Fluorescence Detection X-ray Absorption Spectroscopy
Standard medium
unknown
Observation mode
spectrum
Spectral range type(s)
hard X
Valid spectral range(s)
Min - Max (eV) Sampling (eV) Resolution (eV) Position accuracy (eV) Absorption edge
#1 13351.0 - 13715.0 0.7 Br (K: 14.3256 keV)
Comments
The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the L2 spectrum of a reference Pt metallic foil (first maximum of the derivative set at 13273 eV). The resolution was 2.7eV (FWHM of the elastic peak).
Observation mode
single spot
Date begin
2015-09-03

Versions

Release date
2018-01-31 23:51:01+0000 UTC
Version (Date)
#1 (2018-01-31 23:51:01+0000 UTC, Updated: 2018-11-20 14:23:23+0000 UTC)