Spectrum

NIR-MIR emissivity spectrum of Olivine (sub-µm grains) mixed with KBr powder, Olivine:KBr 9:91 vol%, at 673 K

Title
NIR-MIR emissivity spectrum of Olivine (sub-µm grains) mixed with KBr powder, Olivine:KBr 9:91 vol%, at 673 K
DOI
10.26302/SSHADE/EXPERIMENT_OP_20230206_001
Data reference
Sultana, Robin (2019): NIR-MIR reflectance and emissivity spectra of powdered Olivine (sub-µm grains) mixed with KBr. SSHADE/GhoSST (OSUG Data Center). Dataset/Spectral Data. https://doi.org/10.26302/SSHADE/EXPERIMENT_OP_20230206_001
Publications
Database
Experimentalists
Experiment type
laboratory measurement
Type
thermal emissivity
Intensity unit
no unit
Instrument
Brucker Vertex 80V - UV-Vis-IR reflection
Sample holder
The sample is placed on a stainless steel sample holder, in an external emissivity chamber (DLR-custom-designed) attached to a Bruker Vertex80V FTIR.
Standard medium
air
Observation mode
spectrum
Spectral range type(s)
NIR, MIR
Valid spectral range(s)
Min - Max ($cm^{-1}$) Sampling ($cm^{-1}$) Resolution ($cm^{-1}$) Position accuracy ($cm^{-1}$) Absorption edge
#1 590.0 - 2000.0 1.929 1.0 0.1
Observation geometry
biconical
Observation mode
fixed angles
Incidence angle
0.0°
Emergence angle
30.0°
Azimuth angle
0.0°
Phase angle
30.0°
Resolution observation
5.0°
Illumination
Observation
Comments
some wavelength dependent polarization
Observation mode
single spot
Resolutions
5.0 ${\pm}$ 1.0 $mm$
Experiment
NIR-MIR reflectance and emissivity spectra of powdered Olivine (sub-µm grains) mixed with KBr
Date begin
2019-01-01
Release date
2023-02-08 09:15:15 UTC+0000
Version (Date)
#1 (2023-02-08 09:15:15 UTC+0000, Updated: 2023-02-08 09:15:15 UTC+0000)
History
Date Mode Version Status Comments
2023-02-08 08:58:54 UTC+0000 first import #1 valid version
Analysis
The data were calibrated by dividing the sample radiance by the emitted radiance at the same temperature of a blast furnace slag, taken as blackbody, under the same temperature and observation conditions.
Quality flag
4
Validator(s)